RRRRRRRR RRRRRRRR RR RR eeeee RR RR ee ee aaa RRRRRRRR eeeee a a dme.txt RRRRRRRR ee aaaa RR RR eeeee RR RR RR RR RR RRR MassHunter Workstation Acquisition for 6400 Series Triple Quad LC/MS version B.04.01 Service Pack 7 (SP7) Please read this information carefully. It supersedes the printed material you received with your software. Instructions: Copy executable (*.exe) to the local Hard Disk Drive (HDD) and right mouse click "Run as administrator" especially when installing on Windows 7 64-bit OS. Defect Fixes in SP7: ------------------------------------ SP7 is cumulative and contains all fixes from SP1 to SP6. This includes changes in both Firmware and Software. Firmware version A.00.6.44 automatically starts at the end of SP7 installation. 1) #207859 - Autotune fails to set Mass Axis within specification limit for Wide and Widest Resolution Modes for 6420, 6430, 6460 models Firmware Fixes - version A.00.6.44 ----------------------------------- 1) #189911 - Bug with "estimatedcycletime" for MRM and two uninitialized variables that compute location of settling time. Changes in previous Service Packs: ----------------------------------- • #189911 - Positive-Negative dMRM Performance is Improved with "Dummy" Transitions In B.05 (and B.04) • #198636 - 6490 LogAmp Timing Error on Acq B.05.00 with FW and B.04.01 SP4 with FW 6.36 • #192258 - System is losing sensitivity after it is in standby mode for a while or when spray chamber is opened for cleaning • #195480 - Optimizer does not work for multiple-charged negative ion Optimizer Fix for Presto Number - KEIO0803[2012-Aug03-013245] • #183594 - Fixed the problem of "Adjust" in the checktune report. • #193656 - During a run, divert the LC valve to waste does not function properly • #199399 - Updating Retention Times (with MRM update tool) deletes Positive scan segments • #200779 - Fixed RFPAID's being corrupted by MH eMSEWrite atoms. • #202225 - QuadFreqDip for 1st pass massaxis updated from 100 AMU to 400 AMU for Older Quad Drives. • Optimizer Excel add-in has improved compatibility for Excel 2010 and Windows 7 64-bit OS (helps prevent loss of communication to MS during Excel report generation) • New G6460A high mass peak width calculator improves tune reliability (autotune). • Lens 2 DC stays in correct polarity for ionization mode during autotune (all models). • Other autotune improvements include: Fixed the problem when the data points is only few and the line is flat at peakWidth Axis linearization. Added a little time delay to be able to show the Ion optics ramping curves during Autotune. Fixed the possible software hang in gain calculation. Fixed the problem of "Adjust" in the checktune report. • Generate better quality data for the tune report, it will wait for the expected appropriate data up to 10 times (i.e.: autotune report mass accuracy and abundance identical in checktune). • G6490A ESI with AJST standby conditions are improved. • Chip Cube QQQ standby conditions are improved. • APPI Lamp control is working on G6490A via serial port only (like all other models). • MRM or dMRM acquisition method parameter report print all transitions now. • G1367E Illumination control is now available. Files that are installed with this patch: ----------------------------------------- MassHunter Folder: ------------------ tune\QQQ\G6410B\TunePreferences.xml 4/10/2013 2:56 PM tune\QQQ\G6420A\TunePreferences.xml 4/10/2013 2:56 PM tune\QQQ\G6430A\TunePreferences.xml 4/10/2013 2:56 PM tune\QQQ\G6460A\TunePreferences.xml 4/10/2013 2:56 PM tune\QQQ\G6490A\TunePreferences.xml 4/10/2013 2:56 PM Installation Folder: -------------------- AgtQQQAutotuneAlgos.dll 6/11/2013 11:36 AM Changes from previous Service Packs: ----------------------------------- MassHunter Folder: ------------------ \Configuration\Instruments\LCMS\ENU\agtqqqconfig00.xml 4/4/2011 4:43 PM \Configuration\System\QQQ\app_config.xml 4/4/2011 4:43 PM Instr\default\G6420Adeviceinfo.xml 3/30/2011 5:21 PM Instr\default\G6420AHWParams.xml 4/5/2011 11:28 AM \Report Templates\AcquisitionMethod\Config\192.xml 7/27/2012 12:12 PM \Report Templates\QQQ\RDLTune\CheckTune.rdlc 5/15//2012 9:16 AM \Report Templates\QQQ\RDLTune\CheckTune_MS2.rdlc 5/15//2012 9:16 AM \Report Templates\QQQ\RDLTune\6490A\CheckTune.rdlc 5/15//2012 9:16 AM \Report Templates\QQQ\RDLTune\6490A\CheckTune_MS2.rdlc 5/15//2012 9:16 AM tune\QQQ\G6410B\AgtQQQAutotuneParams.xml 7/8/2011 12:40 AM tune\QQQ\G6420A\AgtQQQAutotuneParams.xml 7/8/2011 12:36 AM tune\QQQ\G6420A\atunes.TUNE.XML 3/30/2011 5.35 PM tune\QQQ\G6430A\AgtQQQAutotuneParams.xml 7/8/2011 12:38 AM tune\QQQ\G6460A\AgtQQQAutotuneParams.xml 7/8/2011 12:39 AM tune\QQQ\G6490A\AgtQQQAutotuneParams.xml 7/11/2011 5:45 PM Installation Folder: -------------------- AgtQQQWorkflowUtility.dll 12/5/2012 3:25 PM AgtOptOptimizer.dll 12/5/2012 2:46 PM agtqqqdevicedriver.dll 12/5/2012 2:42 PM agtqqqdevicedriver.tlb 4/26/2011 11:47 AM AgtQQQTestPeakFinderVC7.dll 4/2/2012 8:44 AM app_ui.exe 1/10/2012 2:14 PM qqqGain.dll 1/10/2012 5:45 PM Microsoft Excel: ---------------- MassHunter Reporting Acq.xlam 7/30/2012 3:04 PM . . o . Agilent Technologies . o . Life Sciences and Chemical Analysis . o o o o . 5301 Stevens Creek Blvd . o . Santa Clara, CA USA . o . 95051 .